Influence of Clipping PTV in Build-Up Region On IMRT Plan Quality and Deliverability

no image available
S Sharma

S Sharma1*, D Manigandan2, P Sahai1 , A Biswas1 , V Subramani1 , S Chander1 , P Julkha1 , G Rath1 , (1) ,,,(2) Fortis Hospital, Mohali, Punjab,

Presentations

SU-E-T-492 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:To study the influence of clipping PTV from body contour on plan quality and deliverability in build-up region for superficial target.

Methods:Five previously treated patients of post-operative carcinoma of parotid were re-planned for IMRT (6MV X-rays, sliding window technique, five fields and 60Gy/30 fractions) using eclipse treatment planning system (TPS) by keeping dose volume constraints and all other parameters constant, only PTV was clipped from body contour by 0mm, 1mm, 2mm and 3mm respectively. Planned fluence was transferred to previously scanned solid water phantom by placing I’matriXX array at 0.5cm depth (2mm slab+3mm inherent). Fluence was delivered by Varian CL2300C/D linac at 99.5cm source to detector distance. Measured fluence was compared with TPS dose plane using 2D gamma evaluation using 3%/3mm DTA criteria. Total MU (monitor unit) required to deliver a plan was also noted. For plan quality, PTV, maximum-dose, minimum-dose, coverage index (CI=PTV volume covered by prescription dose/PTV) and heterogeneity index HI=D5/D95 were analyzed using dose volume histogram (DVH).

Results:The result of gamma function analysis for I’matriXX and TPS were 97.63±1.79%, 97.48±0.99, 98.08±0.89% and 98.01±0.78% at 0.5cm build-up depth for 0, 1, 2 and 3mm PTV clipping, respectively. I’matriXX measured dose was higher compared to TPS. Total MU required for delivering a plan were 552±61, 503±47, 436±24 and 407±22. Maximum-dose to PTV was 6635.80±62.01cGy, 6635.80±40.60cGy, 6608.43±51.07cGy and 6564.20±28.51cGy. Similarly, minimum-dose to PTV was 3306.23±458.56cGy, 3546.57±721.01cGy, 4591.43±298.81cGy and 4861.90±412.40cGy. CI was 0.9347±0.020, 0.9398±0.021, 0.9448±0.022 and 0.9481±0.021. Similarly, HI was 1.089±0.015, 1.084±0.014, 1.078±0.009 and 1.074±0.008 for 0, 1, 2 and 3mm PTV clipping, respectively.

Conclusion:Gamma function analysis resulted in almost similar results. However, I’matriXX was overestimating the dose compared to TPS. MU required to deliver a plan decreases with increase in PTV clipping. CI, PTV minimum-dose and plan homogeneity increases with increase in PTV clipping from skin.