Encrypted login | home

Program Information

Experimental Determination of the Relative Stopping Power of TRUFILL N-BCA Liquid Embolic System in a Clinical Proton Beam


A Mandapaka

A Mandapaka*, A Ghebremedhin , B Patyal , Loma Linda Univ Medical Center, Loma Linda, CA

Presentations

SU-E-T-415 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:Patients who undergo n-BCA glue embolization as part of treatment for AVMs are later referred for proton therapy. Knowing the relative stopping power of the glue accurately allows us to perform accurate dose calculations. In this study we experimentally determine the relative stopping power of an n-BCA mixture in a 126 MeV and 149.6 MeV proton beams.

Methods:: One unit of the TRUFILLâ„¢ n-BCA liquid embolic system consists of 1g unit of n-BCA, 1g unit of Tantalum powder and one 10mL vial of Ethiodized oil. The physician mixed 3:1 Ethiodized oil to n-BCA. Five units (20cc) of the n-BCA liquid embolic glue were prepared and placed in a 6cm x 3cm x3cm Lucite container. The container was placed in front of a water tank in the proton beam path. A diamond detector (active volume 0.004mm3) was used to measure distal edge of depth dose of a modulated 126 MeV proton beam collimated using a 3cm brass aperture. The procedure was repeated with a container carrying the same amount of water placed in front of the water tank. The difference in the depth dose measured with glue and with water was used to determine the relative stopping power of the glue. The same determination was done earlier at 149.6 MeV using a different smaller sample (4cc) of n-BCA.

Results:The relative stopping power of this particular n-BCA mixture was determined to be 1.06 at both 126 MeV and 149.6 MeV. We are working on obtaining the composition data of the n-BCA glue so we can perform Monte Carlo calculations.

Conclusion:Accurate value of the stopping power of the n-BCA glue in the proton beam was determined to be 1.06. It will improve the accuracy of dose calculations in proton radiosurgery procedures on AVM patients with n-BCA embolization.