Tissue Inhomogeneity Corrections in Intra-Operative Radiotherapy

no image available
A Sethi

A Sethi*, B Chinsky , S Gros , A Diak , J Roeske , B Emami , W Small , Loyola University Medical Center, Maywood, IL

Presentations

SU-E-T-756 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: Investigate the impact of tissue inhomogeneities on dose distributions produced by low-energy X-rays in intra-operative radiotherapy (IORT).

Methods: A 50-kV INTRABEAM X-ray device with superficial (Flat and Surface) applicators was commissioned at our institution. For each applicator, percent depth-dose (PDD), dose-profiles (DP) and output factors (OF) were obtained. Calibrated GaFchromic (EBT3) films were used to measure dose distributions in solid water phantom at various depths (2, 5, 10, and 15 mm). All recommended precautions for film-handling, film-exposure and scanning were observed. The effects of tissue inhomogeneities on dose distributions were examined by placing air-cavities and bone and tissue equivalent materials of different density (ρ), atomic number (Z), and thickness (t = 0-4mm) between applicator and film detector. All inhomogeneities were modeled as a cylindrical cavity (diameter 25 mm). Treatment times were calculated to deliver 1Gy dose at 5mm depth. Film results were verified by repeat measurements with a thin-window parallel plate ion-chamber (PTW 34013A) in a water tank.

Results: For a Flat-4cm applicator, the measured dose rate at 5mm depth in solid water was 0.35 Gy/min. Introduction of a cylindrical air-cavity resulted in an increased dose past the inhomogeneity. Compared to tissue equivalent medium, dose enhancement due to 1mm, 2mm, 3mm and 4mm air cavities was 10%, 16%, 24%, and 35% respectively. X-ray attenuation by 2mm thick cortical bone resulted in a significantly large (58%) dose decrease.

Conclusion: IORT dose calculations assume homogeneous tissue equivalent medium. However, soft X-rays are easily affected by non-tissue equivalent materials. The results of this study may be used to estimate and correct IORT dose delivered in the presence of tissue inhomogeneities.