Utility of the Combined Use of Two Types of HDR Sources with the Direction Modulation Brachytherapy (DMBT) Tandem Applicator for Cervical Cancer Treatment


H Safigholi

H Safigholi1*, D Han1,2 , A Soliman1 , A Soleimani Meigooni3 , D Scanderbeg4 , W Song1 , (1) Sunnybrook Research Institute, Sunnybrook Health Sciences Centre, U of T, Toronto, Ontario, (2) University of California, San Diego, La Jolla, California, (3) Comprehensive Cancer Center of Nevada, Las Vegas, Nevada, (4) UCSD Medical Center, La Jolla, CA

Presentations

SU-F-BRA-5 (Sunday, July 12, 2015) 4:00 PM - 6:00 PM Room: Ballroom A


Purpose:To maximize the dose to HRCTV while minimizing dose to the OARs, the combination of two HDR brachytherapy sources, 192-Ir and 169-Yb, used in combination with the recently-proposed novel direction modulated brachytherapy (DMBT) tandem applicator were examined.

Methods:The DMBT tandem, made from nonmagnetic tungsten-alloy rod, with diameter of 5.4mm, has 6 symmetric peripheral holes of 1.3mm diameter. The 0.3mm thick bio-compatible plastic tubing wraps the tandem. MCNPX v.2.6 was used to simulate the mHDR 192-Ir V2 and 4140 HDR 169-Yb sources inside the DMBT applicator. Thought was by combining the higher energy 192-Ir (380keV) and lower energy 169-Yb (92.7keV) sources could create unprecedented level of dose conformality when combined with the high-degree intensity modulation capable DMBT tandem applicator. 3D dose matrices, with 1 mm3 resolution, were imported into an in-house-coded inverse optimization planning system to evaluate plan quality of 19 clinical patient cases. Prescription dose was 15Gy. All plans were normalized to receive the same HRCTV D90.

Results:Generally, the use of dual sources produced better plans than using either of the sources alone, with significantly better performance in some patients. The mean D2cc for bladder, rectum, and sigmoid were 11.65±2.30Gy, 7.47±3.05Gy, and 9.84±2.48Gy for 192-Ir-only, respectively. For 169-Yb-only, they were 11.67±2.26Gy, 7.44±3.02Gy, and 9.83±2.38Gy, respectively. The corresponding data for the dual sources were 11.51±2.24Gy, 7.30±3.00Gy, and 9.68 ±2.39Gy, respectively. The HRCTV D98 and V100 were 16.37±1.86Gy and 97.37±1.92Gy for Ir-192-only, respectively. For 169-Yb-only, they were 16.43±1.86Gy, and 97.51±1.91Gy, respectively. For the dual source, they were 16.42±1.87Gy and 97.47±1.93Gy, respectively.

Conclusion:The plan quality improves, in some cases quite significantly, for when dual 192-Ir and 169-Yb sources are used in combination with highly intensity modulation capable DMBT tandem applicator for image guided cervical cancer brachytherapy.