Development of the Compact Graphite Calorimetry System for the High Energy Photon Beam

no image available
B Kim

B. C. Kim*, I. J. Kim , J. H. Kim, C. Y. Yi , Korea Research Institute of Standards and Science, Daejeon, Republic of Korea

Presentations

SU-E-T-267 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: Graphite calorimeter systems are used for the absolute photon dosimetry. But many electronics are demanded in order to measure the tiny temperature changes. Minimizing the control system is needed to make a portable graphite calorimeter.

Methods: A Domen-type graphite calorimetry system is constructing to measure the absorbed dose of the high energy photon beam. The graphite calorimeter divided into three parts, Core, Jacket, and Shield. In order to measure the temperature rising of the core due to the radiation accurately, the temperatures of the jacket and the shield should be controlled properly. A commercial temperature controller (Model 350, Lake Shore Cryogenics) was used to minimize the size of control system for making a portable graphite calorimetry system at the cost of the measurement uncertainty. The PID control of the jacket is conducted by the software (LabView) and Model 350 maintain the temperature of shield.

Results: Our design value of the heat deposition power in the core is 0.04 mW for the dose rate of 3 Gy/min where the temperature sensitivity of the graphite is 1.4 mK/Gy. While the residuals of the Steinhart-hart equation fitting for the core thermistor were less than 0.1 mK, the temperature resolution of Model 350 is 1 mK. The temperature of the shield was kept within the 5 mK when the room temperature variation was about 0.5 K.

Conclusion: The resolution of Model 350 for the temperature measurement and control is not good enough as the control system for the compact graphite calorimetry system. But The performance of Model 350 is good enough to maintain the temperature of the shield constantly. The Model 350 will be replaced by the AC resistance bridge (Model 372, Lake Shore Cryogenics) for the core temperature measurement and the jacket control.