J Lian

J Lian*, J Matney , E Chao , S Chang , T Zagar , A Wang , B Chera , S Das , E Schreiber , UNC School of Medicine, Chapel Hill, NC

Presentations

SU-C-210-2 (Sunday, July 12, 2015) 1:00 PM - 1:55 PM Room: 210


Purpose:
TomoTherapy treatment has unique challenges in handling intrafractional motion compared to conventional LINAC. This study is aimed to gain a realistic and quantitative understanding of motion impact on TomoTherapy SBRT treatment of lung and prostate cancer patients.

Methods:
A 4D dose engine utilizing GPUs and including motion during treatment was developed for the efficient simulation of TomoTherapy delivered dosimetry. Two clinical CyberKnife lung cases with respiratory motion tracking and two prostate cases with a slower non-periodical organ motion treated by LINAC plus Calypso tracking were used in the study. For each disease site, one selected case has an average motion (6mm); the other has a large motion (10mm for lung and 15mm for prostate). SBRT of lung and prostate cases were re-planned on TomoTherapy with 12 Gyx4 fractions and 7Gyx5 fractions, respectively, all with 95% PTV coverage. Each case was planned with 4 jaw settings: 1) conventional 1cm static, 2) 2.5cm static, 3) 2.5cm dynamic, and 4) 5cm dynamic. The intrafractional rigid motion of the target was applied in the dose calculation of individual fractions of each plan and total dose was accumulated from multiple fractions.

Results:
For 1cm static jaw plans with motions applied, PTV coverage is related to motion type and amplitude. For SBRT patients with average motion (6mm), the PTV coverage remains > 95% for lung case and 74% for prostate case. For cases with large motion, PTV coverage drops to 61% for lung SBRT and 49% for prostate SBRT. Plans with other jaws improve uniformity of moving target, but still suffer from poor PTV coverage (< 70%).

Conclusion:
TomoTherapy lung SBRT is less motion-impacted when average amplitude of respiratory-induced intrafractional motion is present (6mm). When motion is large and/or non-periodic (prostate), all studied plans lead to significantly decreased target coverage in actual delivered dosimetry.