A Dosimetric Evaluation of Boney Anatomy Versus Fiducial Marker Alignment for the Treatment of Prostate Cancer Using Scanned Beam Proton Therapy

no image available
D Freund

D Freund*, X Ding , J Zhang , L Rosen , H Wu , willis knighton medical center, Shreveport, Louisiana

Presentations

SU-E-T-27 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: In prostate proton radiotherapy, three fiducial markers are used for patient daily alignment. However fiducial alignment can change beamline heterogeneity in proton therapy. The purpose of this study is to determine the difference in fiducial and boney anatomy alignment for patient treatment.

Methods and materials: Prostate cancer patients who received proton treatment were included in this study. 3 fiducial markers were implanted before the initial CT. All the patients were re-CT’d every 2 weeks to check the fiducial marker position reproducibility as well as dosimetric consistence of target coverage. In geometry study, re-CT were fused to the initial CT base on the boney anatomy and the average fiducial marker displacement was measured the centers of the fiducials. Dosimetrically, the initial plan was recalculated directly to re-CT image set based on the boney alignment and fiducial alignment to determine the difference from daily treatment. Prostate coverage and hotspots were evaluated using the dose to 98% of the CTV (D98) and dose to 2% (D2), respectively.

Results: The shift from the initial 6 patient CT image sets resulted in an average change in the fiducial location of 5.70 +/- 3 mm. Dosimetric comparison from a single patient revealed that differences from the planned dose resulted from both boney and fiducial alignment. Planned clinical treatment volume coverage resulted in a D98 of 70.44Gy and D2 of 70.84Gy compared to a D98 of 70.13Gy and D2 70.94Gy for boney alignment and a D98 of 70.08Gy and D2 71.18Gy for fiducial alignment respectively.

Conclusion: This study demonstrates that with boney anatomy alignment there is little change to CTV coverage and only slightly worse CTV coverage and hotspot production with fiducial alignment. An increase patient cohort and further investigation is necessary to determine the whether boney alignment can help better control dose heterogeneity.