CT-Based Determination of DIBH Variability and Its Dosimetric Impact On Post-Mastectomy Plus Regional Nodal Radiation Therapy


M Malin

M Malin*, H Kang , K Tatebe , Y Hasan , S Chmura , H Al-Hallaq , The University of Chicago, Chicago, IL

Presentations

SU-F-J-123 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: Breast cancer radiotherapy delivered using voluntary deep inspiration breath-hold (DIBH) requires reproducible breath holds, particularly when matching supraclavicular fields to tangential fields. We studied the impact of variation in DIBHs on CTV and OAR dose metrics by comparing the dose distribution computed on two DIBH CT scans taken at the time of simulation.

Methods: Ten patients receiving 50Gy in 25 fractions to the left chestwall and regional lymph nodes were studied. Two simulation CT scans were taken during separate DIBHs along with a free-breathing (FB) scan. The treatment was planned using one DIBH CT. The dose was recomputed on the other two scans using adaptive planning (Pinnacle 9.10) in which the scans are registered using a cross-correlation algorithm. The chestwall, lymph nodes and OARs were contoured on the scans following the RTOG consensus guidelines. The overall translational and rotational variation between the DIBH scans was used to estimate positional variation between breath-holds. Dose metrics between plans were compared using paired t-tests (p < 0.05) and means and standard deviations were reported.

Results: The registration parameters were sub-millimeter and sub-degree. Although DIBH significantly reduced mean heart dose by 2.4Gy compared to FB (p < 0.01), no significant changes in dose were observed for targets or OARs between the two DIBH scans. Nodal coverage as assessed by V90% was 90%±8% and 89%±8% for supraclavicular and 99%±2% and 97%±22% for IM nodes. Though a significant decrease (10.5%±12.4%) in lung volume in the second DIBH CT was observed, the lung V20Gy was unchanged (14±2% and 14±3%) between the two DIBH scans.

Conclusion: While the lung volume often varied between DIBHs, the CTV and OAR dose metrics were largely unchanged. This indicates that manual DIBH has the potential to provide consistent dose delivery to the chestwall and regional nodes targets when using matched fields.