Encrypted login | home

Program Information

Development and Validation of Dose Calculation for An Open-Source KV Treatment Planning System for Small Animal Radiotherapy


S Prajapati

S Prajapati1*, X Mo2 , B Bednarz2 , M Lawless2 , C Hammer2 , R Flynn3 , D Westerly4 , R Jeraj2 , T Mackie2 , (1) M D Anderson Cancer Center, Houston, TX, (2) University of Wisconsin- Madison, Madison, WI, (3) University of Iowa Hospitals and Clinics, Iowa City, IA, (4) University of Colorado Denver, Aurora, CO,

Presentations

SU-F-T-667 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: An open-source, convolution/superposition based kV-treatment planning system(TPS) was developed for small animal radiotherapy from previously existed in-house MV-TPS. It is flexible and applicable to both step and shoot and helical tomotherapy treatment delivery. For initial commissioning process, the dose calculation from kV-TPS was compared with measurements and Monte Carlo(MC) simulations.

Methods: High resolution, low energy kernels were simulated using EGSnrc user code EDKnrc, which was used as an input in kV-TPS together with MC-simulated x-ray beam spectrum. The Blue Water™ homogeneous phantom (with film inserts) and heterogeneous phantom (with film and TLD inserts) were fabricated. Phantom was placed at 100cm SSD, and was irradiated with 250 kVp beam for 10mins with 1.1cm x 1.1cm open field (at 100cm) created by newly designed binary micro-MLC assembly positioned at 90cm SSD. Gafchromic™ EBT3 film was calibrated in-phantom following AAPM TG-61 guidelines, and were used for measurement at 5 different depths in phantom. Calibrated TLD-100s were obtained from ADCL. EGS and MNCP5 simulation were used to model experimental irradiation set up calculation of dose in phantom.

Results: Using the homogeneous phantom, dose difference between film and kV-TPS was calculated: mean(x)=0.9%; maximum difference(MD)=3.1%; standard deviation(σ)=1.1%. Dose difference between MCNP5 and kV-TPS was: x=1.5%; MD=4.6%; σ=1.9%. Dose difference between EGS and kV-TPS was: x=0.8%; MD=1.9%; σ=0.8%. Using the heterogeneous phantom, dose difference between film and kV-TPS was: x=2.6%; MD=3%; σ=1.1%; and dose difference between TLD and kV-TPS was: x=2.9%; MD=6.4%; σ=2.5%.

Conclusion: The in-house, open-source kV-TPS dose calculation system was comparable within 5% of measurements and MC simulations in both homogeneous and heterogeneous phantoms. The dose calculation system of the kV-TPS is validated as a part of initial commissioning process for small animal radiotherapy. The kV-TPS has the potential for accurate dose calculation for any kV treatment or imaging modalities.



Contact Email: