Optimization of Cyberknife Treatment Planning Using a Dose-Limiting Auto-Shells Method for Brain Metastases


K Yoon

K Yoon1*, B Cho3 , J Kwak3 , D Lee1 , D Kwon2 , S Ahn3 , S Lee3 , C Kim1,2 , S Roh1,2 , Y Cho1,2 , (1) Radiosurgery center, Asan Medical Center, University of Ulsan Collage of Medicine, Seoul, Korea, (2) Department of Neurosurgery, Asan Medical Center, University of Ulsan Collage of Medicine, Seoul, Korea, (3) Department of Radiation oncology, Asan Medical Center, University of Ulsan Collage of Medicine, Seoul, Korea,

Presentations

SU-F-P-60 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:
In pursuit of high precision in target localization and steeper dose fall-off in stereotactic radiosurgery, we investigated the impact of optimizing dose-limiting auto-shell function on the quality of Cyberknife (CK) plans in treating brain metastases (BMs).

Methods:
Nineteen BMs previously treated using CK were selected for this study. The original CK plans (CK_original) had been produced using one to three dose-limiting auto-shells at prescription dose (PD) level and low dose levels of 10 to 30% of PD. In each case, a modified CK plan (CK_modified) was generated using five dose-limiting shells at PD level, intermediate dose level of 50% of PD, and low dose levels with an optimal shell-dilation size based on our experience. In addition, a Gamma Knife plan (GK) was also produced using the original contour set. Thus, a triplet data set of dosimetric parameters was generated and analyzed.

Results:
While no differences in conformity index (mean±SD 1.22 ± 0.1, 1.18 ± 0.1, and 1.24 ± 0.1 in CK_original, CK_modified, and GK, respectively; over all P > 0.05) and tumor coverage (mean±SD 99.5 ± 0.4%, 99.5 ± 0.3%, and 99.4 ± 0.2% in CK_original, CK_modified, and GK, respectively; over all P > 0.05) were observed among the plans, the normal tissue volume receiving 50% of PD was significantly decreased in CK_modified and GK compared with CKoriginal by 1.28 and 1.27-fold, respectively (P < 0.001, each). No significant differences in dose fall-off were observed between CK_modified and GK (P = 0.345).

Conclusion:
By optimizing auto-shell function, a significantly steeper dose fall-off can be achieved in CK system, while maintaining high precision in target localization.