Magnetic Field Dose Response Considerations for a Linac Monitor Chamber


M Reynolds

M Reynolds*, B Fallone , Cross Cancer Institute, Edmonton, AB

Presentations

SU-F-J-147 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: The impact of magnetic fields on the readings of a linac monitor chamber have not yet been investigated. Herein we examine the total dose response as well as any deviations in the beam parameters of flatness and symmetry when a Varian monitor chamber is irradiated within an applied magnetic field. This work has direct application to the development of Linac-MR systems worldwide.

Methods: A Varian monitor chamber was modeled in the Monte Carlo code PENELOPE and irradiated in the presence of a magnetic field with a phase space generated from a model of a Linac-MR prototype system. The magnetic field strength was stepped from 0 to 3.0T in both parallel and perpendicular directions with respect to the normal surface of the phase space. Dose to each of the four regions in the monitor chamber were scored separately for every magnetic field adaptation to evaluate the effect of the magnetic field on flatness and symmetry.

Results: When the magnetic field is perpendicular to the phase space normal we see a change in dose response with a maximal deviation (10-25% depending on the chamber region) near 0.75T. In the direction of electron deflection we expectedly see opposite responses in chamber regions leading to a measured asymmetry. With a magnetic field parallel to the phase space normal we see no measured asymmetries, however there is a monotonic rise in dose response leveling off at about +12% near 2.5T.

Conclusion: Attention must be given to correct for the strength and direction of the magnetic field at the location of the linac monitor chamber in hybrid Linac-MR devices. Elsewise the dose sampled by these chambers may not represent the actual dose expected at isocentre; additionally there may be a need to correct for the symmetry of the beam recorded by the monitor chamber.