Encrypted login | home

Program Information

Optimization of Bladder and Rectal Doses Using a Multi-Lumen Intracavitary Applicator for Gynecological Brachytherapy


S Laoui

S Laoui1*, S Dietrich2 , V Sehgal3 , M Al-Ghazi4 ,University of California, Irvine, Orange, CA.

Presentations

SU-F-T-35 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:Radiation dose delivery for endometrial cancer using HDR techniques is limited by dose to bladder and rectum. A dosimetric study was performed using Varian Capri vaginal brachytherapy applicator to determine the optimal channel configuration which minimizes dose to bladder and rectum, while providing good target coverage.

Methods:A total of 17 patients, 63 plans clinically delivered, and 252 simulated plans using Varian BrachyVision planning system were generated to investigate optimal channel configuration which results in minimum dose to bladder and rectum while providing adequate target coverage. The Capri applicator consists of 13 lumens arranged in two concentric rings, one central lumen and six lumens per ring. Manual dose shaping is invariably required to lower the dose to critical organs. Three-dimensional plans were simulated for 4 channel arrangements, all 13 channels, channel 12 o’clock (close to bladder) and 6 o’clock (close to rectum) deactivated, central channel deactivated, and central channel in addition to 12 o’clock and 6 o’clock deactivated. A relationship between V100, the volume that receives the prescribed dose, and the amount of curie-seconds required to deliver it, was established.

Results:Using all 13 channels results in maximum dose to bladder and rectum. Deactivating central channel in addition to 12 o’clock and 6 o’clock resulted in minimizing bladder and rectum doses but compromised target coverage. The relationship between V100, the volume that receives the prescribed dose, and the curie seconds was found to be linear.

Conclusion:Deactivating channels 12 o’clock and 6 o’clock was shown to be the optimal configuration leading to minimum dose to bladder and rectum without compromising target coverage. The linear relationship between V100 and the curie- seconds can be used as a verification parameter.


Contact Email: