Automated Triplet Beam Orientation Optimization for MRI-Guided Co-60 Radiotherapy


D Nguyen

D Nguyen*, D Thomas , M Cao , D O'Connor , J Lamb , K Sheng , Department of Radiation Oncology, University of California Los Angeles, Los Angeles, CA

Presentations

TH-AB-BRA-2 (Thursday, August 4, 2016) 7:30 AM - 9:30 AM Room: Ballroom A


Purpose:
MRI guided Co-60 provides daily and intrafractional MRI soft tissue imaging for improved target tracking and adaptive radiotherapy. To remedy the low output limitation, the system uses three Co-60 sources at 120° apart, but using all three sources in planning is considerably unintuitive. We automate the beam orientation optimization using column generation, and then solve a novel fluence map optimization (FMO) problem while regularizing the number of MLC segments.

Methods:
Three patients—1 prostate (PRT), 1 lung (LNG), and 1 head-and-neck boost plan (H&NBoost)—were evaluated. The beamlet dose for 180 equally spaced coplanar beams under 0.35 T magnetic field was calculated using Monte Carlo. The 60 triplets were selected utilizing the column generation algorithm. The FMO problem was formulated using an L2-norm minimization with anisotropic total variation (TV) regularization term, which allows for control over the number of MLC segments. Our Fluence Regularized and Optimized Selection of Triplets (FROST) plans were compared against the clinical treatment plans (CLN) produced by an experienced dosimetrist.

Results:
The mean PTV D95, D98, and D99 differ by –0.02%, +0.12%, and +0.44% of the prescription dose between planning methods, showing same PTV dose coverage. The mean PTV homogeneity (D95/D5) was at 0.9360 (FROST) and 0.9356 (CLN). R50 decreased by 0.07 with FROST. On average, FROST reduced Dmax and Dmean of OARs by 6.56% and 5.86% of the prescription dose. The manual CLN planning required iterative trial and error runs which is very time consuming, while FROST required minimal human intervention.

Conclusions:
MRI guided Co-60 therapy needs the output of all sources yet suffers from unintuitive and laborious manual beam selection processes. Automated triplet orientation optimization is shown essential to overcome the difficulty and improves the dosimetry. A novel FMO with regularization provides additional controls over the number of MLC segments and treatment time.


Funding Support, Disclosures, and Conflict of Interest: Varian Medical Systems, NIH grant R01CA188300, NIH grant R43CA183390.