Encrypted login | home

Program Information

Dose Reduction Strategy Using Attenuation-Based Tube Current Modulation Method in CBCT for IGRT

no image available
K Son

K Son1*, H Lee2 , C Kim3 , J Kim4 , S Cho5 , (1) KAIST, Daejeon, (2) KAIST, Daejeon, (3) KAIST, Daejeon, (4) Yonsei Cancer Center, Seoul, (5) KAIST, Daejon,

Presentations

SU-F-J-39 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: To reduce radiation dose to the patients, tube current modulation (TCM) method has been actively used in diagnostic CT systems. However, TCM method has not yet been applied to a kV-CBCT system on a LINAC machine. The purpose of this study is to investigate whether the use of TCM method is desirable in kV-CBCT system for IGRT. We have developed an attenuation–based tube current modulation (a-TCM) method using the prior knowledge of treatment CT image of a patient.

Methods: Patients go through a diagnostic CT scan for RT planning; therefore, using this prior information of CT images, one can estimate the total attenuation of an x-ray through the patient body in a CBCT setting for radiation therapy. We performed a numerical study incorporating major factors into account such as polychromatic x-ray, scatter, noise, and bow-tie filter to demonstrate that a-TCM method can produce equivalent quality of images at reduced imaging radiation doses. Using the CT projector program, 680 projection images of the pediatric XCAT phantom were obtained both in conventional scanning condition, i.e., without modulating the tube current, and in the proposed a-TCM scanning condition. FDK reconstruction algorithm was used for image reconstruction, and the organ dose due to imaging radiation has been calculated in both cases and compared using GATE/Geant4 simulation toolkit.

Results: Reconstructed CT images in the a-TCM method showed similar SSIM values and noise properties to the reference images acquired by the conventional CBCT. In addition, reduction of organ doses ranged from 12% to 27%.

Conclusion: We have successfully demonstrated the feasibility and dosimetric merit of the a-TCM method for kV-CBCT, and envision that it can be a useful option of CBCT scanning that provides patient dose reduction without degrading image quality.


Contact Email: