Encrypted login | home

Program Information

Efficient Dose Calculation Algorithm for FFF IMRT with a Simplified Bivariate Gaussian Source Model


F Li

F Li*, J Park , B Barraclough , B Lu , J Li , C Liu , G Yan , University Florida, Gainesville, FL

Presentations

SU-C-BRC-4 (Sunday, July 31, 2016) 1:00 PM - 1:55 PM Room: Ballroom C


Purpose: To develop an efficient and accurate independent dose calculation algorithm with a simplified analytical source model for the quality assurance and safe delivery of Flattening Filter Free (FFF)-IMRT on an Elekta Versa HD.

Methods: The source model consisted of a point source and a 2D bivariate Gaussian source, respectively modeling the primary photons and the combined effect of head scatter, monitor chamber backscatter and collimator exchange effect. The in-air fluence was firstly calculated by back-projecting the edges of beam defining devices onto the source plane and integrating the visible source distribution. The effect of the rounded MLC leaf end, tongue-and-groove and interleaf transmission was taken into account in the back-projection. The in-air fluence was then modified with a fourth degree polynomial modeling the cone-shaped dose distribution of FFF beams. Planar dose distribution was obtained by convolving the in-air fluence with a dose deposition kernel (DDK) consisting of the sum of three 2D Gaussian functions. The parameters of the source model and the DDK were commissioned using measured in-air output factors (Sc) and cross beam profiles, respectively. A novel method was used to eliminate the volume averaging effect of ion chambers in determining the DDK. Planar dose distributions of five head-and-neck FFF-IMRT plans were calculated and compared against measurements performed with a 2D diode array (MapCHECK™) to validate the accuracy of the algorithm.

Results: The proposed source model predicted Sc for both 6MV and 10MV with an accuracy better than 0.1%. With a stringent gamma criterion (2%/2mm/local difference), the passing rate of the FFF-IMRT dose calculation was 97.2±2.6%.

Conclusion: The removal of the flattening filter represents a simplification of the head structure which allows the use of a simpler source model for very accurate dose calculation. The proposed algorithm offers an effective way to ensure the safe delivery of FFF-IMRT.



Contact Email: