Encrypted login | home

Program Information

A Preliminary Study On Patient Specific VMAT Verification Using a Phosphor-Screen Based Geometric QA System (Raven QA)

no image available
M Lee

M Lee1*, J Wong2 , K Ding2 , B Yi1 , (1) University of Maryland School of Medicine, Baltimore, MD, (2) Johns Hopkins University, Baltimore, MD

Presentations

SU-F-T-287 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: The RavenQA system (LAP Laser, Germany) is a QA device with a phosphor screen detector for performing the QA tasks of TG-142. This study tested if it is feasible to use the system for the patient specific QA of the Volumetric Modulated Arc Therapy (VMAT).

Methods: Water equivalent material (5cm) is attached to the front of the detector plate of the RavenQA for dosimetry purpose. Then the plate is attached to the gantry to synchronize the movement between the detector and the gantry. Since the detector moves together with gantry, The ‘Reset gantry to 0’ function of the Eclipse planning system (Varian, CA) is used to simulate the measurement situation when calculating dose of the detector plate. The same gantry setup is used when delivering the treatment beam for feasibility test purposes. Cumulative dose is acquired for each arc. The optical scatter component of each captured image from the CCD camera is corrected by deconvolving the 2D spatial invariant optical scatter kernel (OSK). We assume that the OSK is a 2D isotropic point spread function with inverse-squared decrease as a function of radius from the center.

Results: Three cases of VMAT plans including head & neck, whole pelvis and abdomen-pelvis are tested. Setup time for measurements was less than 5 minutes. Passing rates of absolute gamma were 99.3, 98.2, 95.9 respectively for 3%/3mm criteria and 96.2, 97.1, 86.4 for 2%/2mm criteria. The abdomen-pelvis field has long treatment fields, 37cm, which are longer than the detector plate (25cm). This plan showed relatively lower passing rate than other plans.

Conclusion: An algorithm for IMRT/VMAT verification using the RavenQA has been developed and tested. The model of spatially invariant OSK works well for deconvolution purpose. It is proved that the RavenQA can be used for the patient specific verification of VMAT.

Funding Support, Disclosures, and Conflict of Interest: This work is funded in part by a Maryland Industrial Partnership Program grant to University of Maryland and to JPLC who owns the Raven technology. John Wong is a co-founder of JPLC.


Contact Email: