Encrypted login | home

Program Information

Evaluation of EBT-XD Radiochromic Films for Verification of SBRT and SRS Treatment Delivery

no image available
J Aubry

J Aubry*, K Zerouali , Centre hospitalier de l'Universite de Montreal, Montreal, Quebec

Presentations

SU-F-T-573 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: To evaluate the accuracy and precision of radiochromic films EBT-XD for quality control of stereotaxic body radiation therapy (SBRT) and stereotactic radiosurgery (SRS) plan delivery.

Methods: A film cut in 3x10 cm2 strips was irradiated from 0 to 20 Gy in increments of 1 to 1.5 Gy using a 15 MeV electron beam. Triple-channel film calibration was completed 24 hours later by scanning the film strips on an Epson 10000XL scanner using a well-defined protocol. Several dose measurements of increasing complexity were subsequently performed with Varian iX accelerators. Pieces of films were first irradiated in a solid water phantom with 6 MV photon beams and a static gantry to doses spanning the calibration range, either in a single field or multiple fields setup. High dose (>15 Gy per fraction) IMRT plans were then measured. Finally films were irradiated with volumetric modulated arc therapy (VMAT) plans of lung and spinal lesions with prescribed doses per fraction between 8 and 20 Gy. The dose measured with the films was compared to the calculated dose from the Eclipse planning system using the Anisotropic Analytical Algorithm (AAA).

Results: 77 dose measurements were compared to either ion chamber measurements or dose calculations (reference). The average dose difference between film measurements and reference was 0.7 % and the standard deviation was 1.3%. The maximum and minimum dose differences were +3.5% and -2% in the 4 Gy to 20 Gy range. Measured dose profiles of lung and vertebra treatment plans agreed very well with the calculations.

Conclusion: EBT-XD films are a useful dosimeter for quality control of SBRT and SRS plan delivery. The measurement of a full 2D dose plane with high spatial resolution and acceptable dose accuracy make it an advantageous choice compared to other detectors such as ion chambers or diodes.



Contact Email: