Encrypted login | home

Program Information

A Post-Mastectomy Radiation Therapy Dose Distribution Study Using Nanodots and Films


X Qian

X Qian*, K Vaidya , L Puckett , S Souri , Y Chen , Y Cao , A Jamshidi , L Lee , E Klein , F Diaz , Long Island Jewish Medical Center, Lake Success, NY

Presentations

SU-F-T-323 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:In post-mastectomy radiation therapy (RT), skin dose must be accurately estimated to assess skin reactions such as erythema, desquamation and necrosis. Planning systems cannot always provide accurate dosimetry for target volumes distal to skin. Therefore, in-vivo dosimetry is necessary. A female anthropomorphic phantom was used with optically stimulated luminescence dosimeters (nanoDots) to measure dose to chest wall skin. In addition, EBT2 films was employed to measure dose to left lung and heart in post-mastectomy RT.

Methods:Films and nanoDots were calibrated under full buildup conditions at 100cm SAD for 6MV photons. Five pieces of films were placed between slabs of Rando phantom to assess dose to left lung and heart. Two layers of 0.5cm thick bolus were used to cover the whole left chest. Six pairs of nanoDots were placed at medical and lateral aspects on the bolus surface, between the 0.5cm bolus layers, and under the bolus. Three control nanoDots were placed on chest wall to quantify imaging dose. The phantom was CT scanned with all dosimeters in place, and treatment planning was performed with tangential fields (200cGy). All dosimeters were contoured on CT and dose was extracted. NanoDots were read using nanoDot reader and films were scanned using film scanner. The measured and calculated doses were tabulated.

Results:Dose to 12 nanoDots were evaluated. Dose variance for surface nanoDots were +3.8%, +2.7%, -5% and -9.8%. Those at lateral positions, with greater beam obliquity had larger variance than the medial positions. A similar trend was observed for other nanoDots (Table1). Point doses from films for heart and the left lung were 112.7cGy and 108.7cGy, with +10.2% and +9.04% deviation from calculated values, respectively.

Conclusion:Dosimetry provided by the advanced planning system was verified using NanoDots and films. Both nanoDots and films provided good estimation of dose distribution in post-mastectomy RT.


Contact Email: