Encrypted login | home

Program Information

Transit Dose Comparisons for Co-60 and Ir-192 HDR Sources


J Vijande

V Gimenez-Alventosa1 , J Vijande2*, F Ballester1 , J Perez-Calatayud3 , (1) University of Valencia, Burjassot, Spain, (2) University of Valencia and IFIC(CSIC-UV), Burjassot, Spain, (3) Hospital Clinica Benidorm, Benidorm, and Hospital Universitari i Politecnic La Fe, Valencia, Spain

Presentations

SU-F-T-13 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: The purpose of this study is to compare the transit dose due to the movement of high dose rate (HDR) Ir-192 and Co-60 sources along the transfer tube. This is performed by evaluating air-kerma differences in the vicinity of the transfer tube when both sources are moved with the same velocity from a HDR brachytherapy afterloader into a patient.

Methods: Monte Carlo simulations have been performed using PENELOPE2014. mHDR-v2 and Flexisource sources have been considered. Collisional kerma has been scored. The sources were simulated within a plastic catheter located in an infinite air phantom. The movement of the seed was included by displacing their positions along the connecting catheter from z=-75 cm to z=+75 cm and combining them. Backscatter from the afterloader and the patient was not considered. Since modern afterloaders like Flexitron (Elekta) or Saginova (Bebig) are able to use equally Ir-192 and Co-60 sources it was assumed that both sources are displaced with equal speed. Typical content activity values were provided by the manufacturer (460 GBq for Ir-192 and 75 GBq for Co-60).

Results: 2D distributions were obtained with type-A uncertainties (k=2) less than 0.01%. From those, the air kerma ratio Co-60/Ir-192 was evaluated weighted by their corresponding activities. It was found that it varies slowly with distance (less than 10% variation) but strongly in time due to the shorter half-life of the Ir-192 (73.83 days). The maximum ratio is located close to the catheter with a value of 0.57 when both sources are installed by the manufacturer, while increasing up to 1.25 at the end of the recommended working life (90 days) of the Ir-192 source.

Conclusion: Air-kerma ratios are almost constant (0.51-0.57) in the vicinity of the source. Nevertheless, air-kerma ratios increase rapidly whenever the Ir-192 is approaching the end of its life.



Contact Email: