Mathematical Estimation and Experimental Measurement of Patient Free-In-Air Skin Entrance Exposure During a Panoramic Dental X-Ray Procedure

no image available
A Errico

A Errico*, R Behrman , B Li , Boston University Medical Center, Boston, MA

Presentations

TU-FG-209-9 (Tuesday, August 2, 2016) 1:45 PM - 3:45 PM Room: 209


Purpose: To develop a simple mathematical model for estimating the patient free-in-air skin entrance exposure (SEE) during a panoramic dental x-ray that does not require the use of a head phantom. This eliminates issues associated with phantom centering and the mounting of a detector on the phantom for routine QC testing.

Methods: We used a Sirona Orthophos XG panoramic radiographic unit and a Radcal Accu-Gold system for this study. A solid state detector was attached over the slit of the Orthophos’ sensor with the help of a custom-built jig. A single measurement of the free-in-air exposure at this position was taken over a full panoramic scan. A mathematical model for estimating the SEE was developed based upon this measurement, the system geometry, x-ray field beam width, and x-ray sweep angle.

To validate the model, patient geometry was simulated by a 16 cm diameter PMMA CTDI phantom centered at the machine’s isocenter. Measurements taken on the phantom’s surface were made using a solid state detector with lead backing, an ion chamber, and the ion chamber with the phantom wrapped in lead to mitigate backscatter. Measurements were taken near the start position of the tube and at 90 degrees from the start position.

Results: Using the solid state detector, the average SEE was 23.5+/-0.02 mR and 55.5+/-0.08 mR at 64 kVp and 73 kVp, respectively. With the lead-wrapping, the measurements from the ion chamber matched those of the solid state detector to within 0.1%. Preliminary results gave the difference between the mathematical model and the phantom measurements to be approximately 5% at both kVps.

Conclusion: Reasonable estimates of patient SEE for panoramic dental radiography can be made using a simple mathematical model without the need for a head phantom.