Comparative Study Between Two Normalization Prescriptions for Accelerated Partial Breast Irradiation: A Dosimetric Study

no image available
P Agarwal

P Agarwal*, D Sharma , A Gandhi , A Binjola , V Subramani , S Chander , All India Institute of Medical Sciences, New Delhi, New Delhi, Delhi

Presentations

SU-F-P-46 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:To compare the Accelerated Partial Breast Irradiation (APBI) plan with the normalized basal dose points and 5mm box prescription.

Methods:Five patients of APBI were planned twice in Oncentra Master planning TPS (Version 4.3) using TG-43 algorithm. The number of catheters for all the patients was 10 to 16 and implant plane 2 to 3. For planning all catheters were reconstructed. Source loading was done as per HR-CTV contoured. The HR-CTV volume range was from 75cc to 182cc. Plans were normalized in two methods. First all plans were normalized on Basal dose points (PlanA) and second all the plan were normalized on 5mm box (PlanB). The prescription dose (PD) was 35Gy in 10 fractions. All the plans were completely based on normalization and without optimization. Plan evaluation was based on certain parameters coverage Index (CI), dose homogeneity index (DHI), conformity index (COIN), over dose volume index (OI).

Results:The average and median of CI for planA was 0.835 and 0.8154, for planB 0.82 and 0.799 respectively. The median and average of DHI for planA was 0.66 and 0.6062, for planB 0.67 and 0.62 respectively. The range of COIN for planA and planB was from 0.58 to 0.65 respectively. The range of OI was from 0.083 to 0.169 for planA and planB. The treatment time in planA was in average 1.13 times more than planB as V150% of HR_CTV in planA was 4-6% more. The ipsilateral lung was getting 30% of PD which was 0.6% to 3.5%.

Conclusion:Treatment Planning should be individualized based on implants characteristics. Planning with prescription to basal dose points should be preferred to 5mm box prescription, in order to achieve better DHI and less treatment time.