Toward Heavy Ion Computed Tomography with Carbon Ions: A Monte Carlo Study

no image available
D Shrestha

D Shrestha*, N Qin , Y Zhang , X Jia , J Wang , UT Southwestern Medical Center, Dallas, TX

Presentations

SU-G-IeP2-13 (Sunday, July 31, 2016) 4:30 PM - 5:00 PM Room: ePoster Theater


Purpose:In the present Monte Carlo study, we investigated the use of Carbon ions for computed tomography (CT) with a relatively low imaging dose. This will enable us to avoid any conversion of X-ray CT numbers to the relative stopping power (or relative electron density) values and the associated uncertainties in Carbon dose calculation.

Methods:In the first stage, we studied the propagation of Carbon nuclei through a water phantom using the Geant4 specially to understand their lateral displacement inside the phantom. In the second stage, we used our GPU-based Monte Carlo code, which has been cross validated against Geant4, to create the 2D map of the water equivalent path length (WEPL) inside a human head size phantom by acquiring 240 projections each 1.5 ° apart. Subsequently the 3D relative electron density map of the phantom was reconstructed from the 2D WEPL map using the Algebraic Reconstruction Technique (ART) coupled with total variation (TV) minimization

Results:A high quality image of the relative electron density inside the phantom was reconstructed by ARTTV. The mean relative error between the reconstructed image for low contrast object (PMMA) was about 1.74%. The delivered dose per scan at the center of the phantom was about 0.1 Gy.

Conclusion:We have been able to obtain a 3D map of the electron density using a human head size phantom while keeping the delivered dose to relatively low value. Using the GPU capabilities of our simulation engine, we believe that a real time CT with Carbon ions could be a reality in future.