Encrypted login | home

Program Information

Reproducibility of Tumor Motion Probability Distribution Function in Stereotactic Body Radiation Therapy of Lung Using Real-Time Tumor-Tracking Radiotherapy System

no image available
T Shiinoki

T Shiinoki1*, S Kawamura2 , H Hanazawa1 , S Park1 , T Uehara2 , Y Yuasa2 , M Koike2 ,T Takahashi1 ,K Shibuya1 , (1) Yamaguchi University, Ube, Yamaguchi, (2) Yamaguchi University Hospital, Ube, Yamaguchi.

Presentations

SU-E-J-182 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:We aim to achieve new four-dimensional radiotherapy (4DRT) using the next generation real-time tumor-tracking (RTRT) system and flattening-filter-free techniques. To achieve new 4DRT, it is necessary to understand the respiratory motion of tumor. The purposes of this study were: 1.To develop the respiratory motion analysis tool using log files. 2.To evaluate the reproducibility of tumor motion probability distribution function (PDF) during stereotactic body RT (SBRT) of lung tumor.

Methods:Seven patients having fiducial markers closely implanted to the lung tumor were enrolled in this study. The positions of fiducial markers were measured using the RTRT system (Mitsubishi Electronics Co., Japan) and recorded as two types of log files during the course of SBRT. For each patients, tumor motion range and tumor motion PDFs in left-right (LR), anterior-posterior (AP) and superior-inferior (SI) directions were calculated using log files of all beams per fraction (PDFn). Fractional PDF reproducibility (Rn) was calculated as Kullback-Leibler (KL) divergence between PDF1 and PDFn of tumor motion. The mean of Rn (Rm) was calculated for each patient and correlated to the patient’s mean tumor motion range (Am). The change of Rm during the course of SBRT was also evluated. These analyses were performed using in-house developed software.

Results:The Rm were 0.19 (0.07-0.30), 0.14 (0.07-0.32) and 0.16 (0.09-0.28) in LR, AP and SI directions, respectively. The Am were 5.11 mm (2.58-9.99 mm), 7.81 mm (2.87-15.57 mm) and 11.26 mm (3.80-21.27 mm) in LR, AP and SI directions, respectively. The PDF reproducibility decreased as the tumor motion range increased in AP and SI direction. That decreased slightly through the course of RT in SI direction.

Conclusion:We developed the respiratory motion analysis tool for 4DRT using log files and quantified the range and reproducibility of respiratory motion for lung tumors.




Contact Email: