Encrypted login | home

Program Information

Initial Characterization of a Novel 2D Computed Radiography (CR) Dosimeter for SBRT

no image available
W Crijns

W Crijns1, 2*, S Ramaekers1, P Leblans3, G Defraene1 , F Maes4, 5 , K Haustermans1 , T Depuydt1 , (1) Department of Oncology, Laboratory of Experimental Radiotherapy, KU Leuven, Leuven, Belgium, (2) Medical Imaging Research Center, KU Leuven, Leuven, Belgium, (3) Agfa HealthCare NV, Imaging, R&D Consumables, Mortsel, Belgium, (4) Department of Electrical Engineering (ESAT) - PSI, KU Leuven, Leuven, Belgium, (5) iMinds - Medical IT department, KU Leuven, Leuven, Belgium

Presentations

SU-E-T-497 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: For 2D, sub-mm resolution dose measurements Gafchromic™ EBT film is currently a standard in radiotherapy; mainly because of its energy independence and water equivalence. However, EBT film is disposable. Therefore, the dosimetric and uniformity characteristics needs to be estimated from a second, possibly different, (calibration) film. Moreover, EBT has a post-irradiation coloration and a non-linear dose dependence with saturation, which limits the applicable time interval and dose range.

CR technology forms an interesting alternative for EBT. Dose dependent CR-plates have a sub-mm resolution, and additionally a linear dose dependence over decades of dose. But, CR has an inherent signal fading and energy dependence. Here, for the first time, a radiotherapy 2D CR prototype was characterized for an extended dose range (up to 35Gy), signal fading, and basic energy dependence.

Methods: The prototype was irradiated with a standard 10x10cm 6MV photon beam, and scanned with a commercial CR 15-X(R) scanner. The time between the start of irradiation and scanning (T_Scan) was monitored. The linearity of the dose response was evaluated between 0 and 35Gy using a fixed T_Scan of 4min, if possible (i.e. ≤10Gy). Next, the signal fading was characterized for a T_Scan-range of 4 to 20min. The energy dependence was assessed by a comparison of out-of-field measurements of CR, EBT, and TLD.

Results: The radiotherapy CR prototype has a linear response over the complete SBRT dose range (0-35Gy). The prototype had a small (5%), but linear signal fading over the time interval of interest (4-20min). Out-of-field, the prototype has a 8% over response due to an increased amount of low energy photons. The impact of the over response on intensity modulated radiotherapy remains to be evaluated.

Conclusion: CR technology is promising for SBRT dose measurements up to 35Gy. It is a reusable linear alternative for film dosimetry.

Funding Support, Disclosures, and Conflict of Interest: Research partly supported by Agfa HealthCare NV (Mortsel, Belgium)


Contact Email: