Utility of the Two Candidate 192-Ir and 169-Yb HDR Sources for Use with a Novel Direction Modulated Brachytherapy Tandem Applicator for Cervical Cancer Treatment


H Safigholi

H Safigholi1*, D Han1,2 , A Soliman1 , A Soleimani Meigooni3 , W Song1 , (1) Sunnybrook Research Institute, Sunnybrook Health Sciences Centre, U of T, Toronto, Ontario, (2) University of California, San Diego, La Jolla, California, (3) Comprehensive Cancer Center of Nevada, Las Vegas, Nevada

Presentations

SU-E-T-787 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:A novel tungsten alloy shielded, MRI-compatible, direction modulated brachytherapy (DMBT) concept tandem applicator, which enables unprecedented intensity modulation, was used to evaluate treatment plan quality improvement over a conventional tandem. The utility of the 192-Ir and 169-Yb HDR sources, for use with the DMBT applicator, was evaluated.

Methods:The total diameter of the DMBT tandem applicator is 6.0 mm, which consists of 5.4-mm diameter tungsten alloy and 0.3 mm thick plastic sheath. The tandem has 6 symmetric peripheral 1.3-mm diameter grooves for the source to travel. MCNPX v.2.6 was used to simulate the 192-Ir and 169-Yb sources inside the DMBT applicator. First, TG-43 source parameters were evaluated. Second, 3D dose matrix with 1 mm3 resolution were imported into an in-house-coded inverse optimization treatment planning program to obtain optimal plans for 19 clinical cases. All plans were compared with the standard tandem and ring plans. Prescription dose was 15.0 Gy. All plans were normalized to receive the same HRCTV D90.

Results:Generally, the DMBT tandem (and ring) plans were better than the conventional tandem and ring plans for 192-Ir and 169-Yb HDR sources. The mean data of D2cc for bladder, rectum, and sigmoid were 11.65±2.30 Gy, 7.47±3.05 Gy, and 9.84±2.48 Gy for Ir-192 DMBT tandem, respectively. These data for Yb-169 were 11.67±2.26 Gy, 7.44±3.02 Gy, and 9.83±2.38 Gy, respectively. The HR-CTV D98 and V100 were 16.37±1.86 Gy and 97.37 ± 1.92 Gy for Ir-192 DMBT, respectively. The corresponding values for Yb-169 were 16.43±1.86 Gy, and 97.51 ± 1.91 Gy. Plans with the 169-Yb source generally produced more favorable results where V100 increased by 13.65% while D2cc across all OARs reduced by 0.54% compared with the 192-Ir plans.

Conclusion:For the DMBT tandem applicator, 169-Yb source seems to produce more directional beams resulting in increased intensity modulation capacity, thus resulting in more conformal plans