Encrypted login | home

Program Information

Analysis of Total Imaging Uncertainty in Respiratory-Gated Radiotherapy

no image available
J Suzuki

J Suzuki1*, T Okuda1 , S Sakaino2 , N Yokota2 , (1) Toyota memorial hospital, Toyota, Aichi, (2) Suzukake central hospital, Hamamatsu, Shizuoka

Presentations

SU-E-J-159 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:In respiratory-gated radiotherapy, the gating phase during treatment delivery needs to coincide with the corresponding phase determined during the treatment plan. However, because radiotherapy is performed based on the image obtained for the treatment plan, the time delay, motion artifact, volume effect, and resolution in the images are uncertain. Thus, imaging uncertainty is the most basic factor that affects the localization accuracy. Therefore, these uncertainties should be analyzed. This study aims to analyze the total imaging uncertainty in respiratory-gated radiotherapy.

Methods:Two factors of imaging uncertainties related to respiratory-gated radiotherapy were analyzed. First, CT image was used to determine the target volume and 4D treatment planning for the Varian Real-time Position Management (RPM) system. Second, an X-ray image was acquired for image-guided radiotherapy (IGRT) for the BrainLAB ExacTrac system. These factors were measured using a respiratory gating phantom. The conditions applied during phantom operation were as follows: respiratory wave form, sine curve; respiratory cycle, 4 s; phantom target motion amplitude, 10, 20, and 29 mm (which is maximum phantom longitudinal motion). The target and cylindrical marker implanted in the phantom coverage of the CT images was measured and compared with the theoretically calculated coverage from the phantom motion. The theoretical position of the cylindrical marker implanted in the phantom was compared with that acquired from the X-ray image. The total imaging uncertainty was analyzed from these two factors.

Results:In the CT image, the uncertainty between the target and cylindrical marker’s actual coverage and the coverage of CT images was 1.19 mm and 2.50mm, respectively. In the X-ray image, the uncertainty was 0.39 mm. The total imaging uncertainty from the two factors was 1.62mm.

Conclusion:The total imaging uncertainty in respiratory-gated radiotherapy was clinically acceptable. However, an internal margin should be added to account for the total imaging uncertainty.


Contact Email: