Encrypted login | home

Program Information

Investigation of the CyberKnife MultiPlan Monte Carlo Dose Calculation Using EBT3 Film Absolute Dosimetry for Delivery in a Heterogeneous Thorax Phantom


M Lamberto

M Lamberto*, H Chen , K Huang , F Mourtada , Christiana Care Hospital, Helen F. Graham Cancer Center, Newark, DE

Presentations

SU-E-T-519 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: To characterize the Cyberknife (CK) robotic system’s dosimetric accuracy of the delivery of MultiPlan’s Monte Carlo dose calculations using EBT3 radiochromic film inserted in a thorax phantom.

Methods: The CIRS XSight Lung Tracking (XLT) Phantom (model 10823) was used in this study with custom cut EBT3 film inserted in the horizontal (coronal) plane inside the lung tissue equivalent phantom. CK MultiPlan v3.5.3 with Monte Carlo dose calculation algorithm (1.5 mm grid size, 2% statistical uncertainty) was used to calculate a clinical plan for a 25-mm lung tumor lesion, as contoured by the physician, and then imported onto the XLT phantom CT. Using the same film batch, the net OD to dose calibration curve was obtained using CK with the 60 mm fixed cone by delivering 0-800 cGy. The test films (n=3) were irradiated using 325 cGy to the prescription point. Films were scanned 48 hours after irradiation using an Epson v700 scanner (48 bits color scan, extracted red channel only, 96 dpi). Percent absolute dose and relative isodose distribution difference relative to the planned dose were quantified using an in-house QA software program. Multiplan Monte Carlo dose calculation was validated using RCF dosimetry (EBT3) and gamma index criteria of 3%/3mm and 2%/2mm for absolute dose and relative isodose distribution measurement comparisons.

Results: EBT3 film measurements of the patient plans calculated with Monte Carlo in MultiPlan resulted in an absolute dose passing rate of 99.6±0.4% for the Gamma Index of 3%/3mm, 10% dose threshold, and 95.6±4.4% for 2%/2mm, 10% threshold criteria. The measured central axis absolute dose was within 1.2% (329.0±2.5 cGy) of the Monte Carlo planned dose (325.0±6.5 cGy) for that same point.

Conclusion: MultiPlan’s Monte Carlo dose calculation was validated using the EBT3 film absolute dosimetry for delivery in a heterogeneous thorax phantom.



Contact Email: