Encrypted login | home

Program Information

Output Factor Calculation for a Uniform Scanning Proton Therapy System

no image available
D Hecksel

D Hecksel*, N Stauffer , G DeFillippo , J Edwards , M Pankuch , Northwestern Medicine Chicago Proton Center, Warrenville, IL

Presentations

SU-E-T-595 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: To develop an analytical model that predicts dose output factors for patient treatment fields in a uniform scanning proton therapy system.

Methods: A model to predict output factors for patient specific treatment fields was produced based on the methods developed by Kooy et al. (2003). The Kooy model predicts the output factor based on the ratio of the entrance dose under calibration conditions to that for a given range and modulation corrected for the inverse square effect. Field specific output factors were plotted as a function of a single parameter r, where r = (Range-Modulation)/Modulation. The model targeted user range 1 sub-span 3 through user range 2 sup-span 2 of the IBA uniform scanning proton therapy system. The data set included points measured using the 10 cm and 18 cm snout sizes to eliminate stem effects on the monitor unit chambers. The data was fit using equation 15 from Kooy et al. (2003), and the resulting model was tested against measurements that were not included in the original data set.

Results: For the range and sub span investigated, 120 data points were tested against the model prediction. The model predicted the output factor within 2% for 96% of the points tested and within 2.5% for 99% of the points tested. All points were within 3% of the predicted values.

Conclusion: Monitor units for patient treatment fields with proton ranges that fall within the tested interval can be predicted using a model based on the methods developed at MGH. With further evaluation, it will be possible to model all user ranges and sub-spans of the IBA system. Further testing is also needed to predict output factors using a 25 cm snout which introduces variable scanning pattern sizes and stem effects on the monitor unit chambers.


Contact Email: