Question 1: Which of the following statements about homogeneity testing using the manufacturer's procedure is not correct?
Reference:American College of Radiology 2015 MRI Quality Control Manual
Choice A:Results using the vendor’s testing procedure are most likely to be accepted by the vendor’s field service engineer.
Choice B:ACR accreditation requires homogeneity to be measured using the manufacturer’s testing procedure
Choice C:Homogeneity specified by the manufacturer are established using the manufacturer’s testing procedure.
Choice D:The manufacturer’s testing procedure may require tools that are not readily available to the physicist.
Question 2: According to the American College of Radiology 2015 MRI Quality Control Manual, which of the following statements concerning suggested homogeneity performance criteria is not true?
Reference:American College of Radiology 2015 MRI Quality Control Manual
Choice A:Homogeneity criteria should be included in the Annual Performance Evaluation.
Choice B:Homogeneity data are particular to the diametrical spherical volume used.
Choice C:Obtained homogeneity values should agree closely with the manufacturer’s specifications
Choice D:Results will differ based on whether ptp or rms is measured.
Choice E:No specific values for homogeneity are indicated.
Question 3: Which method can be used to determine homogeneity in multiple diameters of spherical volume (DSV) in a single uniform fluid-filled phantom?
Reference:American College of Radiology 2015 MRI Quality Control Manual and Chen HH et al, Routine testing of magnetic field homogeneity on clinical MRI systems. Med Phys 2006, 33(11): 4299-4306.
Choice A:Spectral peak
Choice B:Bandwidth difference
Choice C:Phase difference
Question 4: True or false: Both peak-to-peak and rms B0 inhomogeneity can be measured with pre-scan spectral peak, bandwidth difference, and phase difference methods.
Reference:American College of Radiology 2015 MRI Quality Control Manual
Choice A:True
Choice B:False
Question 5: The bandwidth difference approach to B0 homogeneity testing measures the field homogeneity in the direction of which axis?
Reference:American College of Radiology 2015 MRI Quality Control Manual
Choice A:Frequency encoding
Choice B:Phase encoding
Choice C:Slice selection
Question 6: True or false: decreasing the difference between echo times will decrease the number of phase wraps in a phase difference image.
Reference:American College of Radiology 2015 MRI Quality Control Manual
Choice A:True
Choice B:False