Usefulness of CT-MR Fusion in Radiotherapy Planning for Prostate Cancer Patient with Bilateral Hip Replacements

no image available
R He

R. He1,2*, P. Kumar2, Y. Hu1 , J. Suggs1, C. Yang1 , Shankar Giri1,2 (1) University of Mississippi Med. Center, Jackson, MS, (2) VA Medical Center at Jackson, Mississippi

Presentations

SU-E-T-504 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose: Target localization of prostate for Intensity Modulated Radiation Therapy (IMRT) in patients with bilateral hip replacements is difficult due to artifacts in Computed Tomography (CT) images generated from the prostheses high Z materials. In this study, Magnetic Resonance (MR) images fused with CT images are tested as a solution.
Methods: CT images of 2.5 mm slice thickness were acquired on a GE Lightspeed scanner with a flat-topped couch for a prostate cancer patient with bilateral hip replacements. T2 weighted images of 5 mm separation were acquired on a MR Scanner. After the MR-CT registration on a radiotherapy treatment planning system (Eclipse,Varian), the target volumes were defined by the radiation oncologists on MR images and then transferred to CT images for planning and dose calculation. The CT Hounsfield Units (HU) was reassigned to zero (as water) for artifacts. The Varian flat panel treatment couch was modeled for dose calculation accuracy with heterogeneity correction. A Volume Matrix Arc Therapy (VMAT) and a seven-field IMRT plans were generated, each avoiding any beam transversing the prostheses; the two plans were compared. The superior VMAT plan was used for treating the patient. In-vivo dosimetry was performed using MOSFET (Best Canada) placed in a surgical tube inserted into the patient rectum during therapy. The measured dose was compared with planned dose for MOSFET location.
Results: The registration of MR-CT images and the agreement of target volumes were confirmed by three physicians. VMAT plan was deemed superior to IMRT based on dose to critical nearby structures and overall conformality of target dosing. In-vivo measured dose compared with calculated dose was -4.5% which was likely due to attenuation of the surgical tube surrounding MOSFET.
Conclusion: When artifacts are present on planning CT due to bilateral hip prostheses, MR-CT image fusion is a feasible solution for target delineation.