Online Only Report No. OR02 - A protocol for the determination of absorbed dose from high-energy photon and electron beams. (1983) Category: Online Only Reports A protocol is given to provide radiological physicists with an accurate means of measuring radiation doses on phantoms. Plastics, as well as water, are considered as phantoms. A new quantity, the cavity-gas calibration factor, is introduced. Medical Physics, 10, 741-771 (1983) https://doi.org/10.1118/1.595446 Altmetrics for this report Radiation Therapy Task Group #21 2542 KB Robert Schulz, Peter Almond, John Cunningham, J. Garrett Holt, Robert Loevinger, Nagalingam Suntharalingam, Kenneth A. Wright, Ravinder Nath, Geoffrey Lempert |