Modeling the Dosimetric Effects of Volumetric and Layer-Based Repainting Strategies in Spot Scanning Proton Treatment Plans


J Johnson

J E Johnson*, C Beltran , M G Herman , J J Kruse , Mayo Clinic, Rochester, MN

Presentations

SU-D-BRE-6 Sunday 2:05PM - 3:00PM Room: Ballroom E

Purpose: To compare multiple repainting techniques as strategies for mitigating the interplay effect in free-breathing, spot scanning proton plans.

Methods: An analytic routine modeled three-dimensional dose distributions of pencil-beam proton plans delivered to a moving target. The interplay effect was studied in subsequent calculations by modeling proton delivery from a clinical synchrotron based spot scanning system and respiratory target motion, patterned from surrogate breathing traces from clinical 4DCT scans and normalized to nominal 0.5 and 1 cm amplitudes. Two distinct repainting strategies were modeled. In idealized volumetric repainting, the plan is divided up and delivered multiple times successively, with each instance only delivering a fraction of the total MU. Maximum-MU repainting involves delivering a fixed number of MU per spot and repeating a given energy layer until the prescribed MU are reached. For each of 13 patient breathing traces, the dose was computed for up to four volumetric repaints and an array of maximum-MU values. Delivery strategies were inter-compared based on target coverage, dose homogeneity, and delivery time.

Results: Increasing levels of repainting generally improved plan quality and reduced dosimetric variability at the expense of longer delivery time. Motion orthogonal to the scan direction yielded substantially greater dose deviations than motion parallel to the scan direction. For a fixed delivery time, maximum-MU repainting was most effective relative to idealized volumetric repainting at small maximum-MU values. For 1 cm amplitude motion orthogonal to the scan direction, the average homogeneity metric (D5 – D95)[%] of 23.4% was reduced to 7.6% with a 168 s delivery using volumetric repainting compared with 8.7% in 157.2 s for maximum-MU repainting. The associated static target homogeneity metric was 2.5%.

Conclusion: Maximum-MU repainting can provide a reasonably effective alternative to volumetric repainting for systems without the capability to switch energies rapidly.